Pumping and pressure reading
Pagina 1 di 6
The pressure in the chambers should not exceed these values during the following operations:Operation | Analysis chamber | Preparation chamber - bottom | Preparation chamber - top | Load lock | Electrochemical cell |
sample transfer | 2×10-8 mbar | 5×10-7 mbar | 5×10-7 mbar | 1×10-6 mbar | 1×10-5 mbar |
ion sputtering | 2×10-6 mbar | 2×10-6 mbar | – | – | – |
gas exposure | 1×10-6 mbar | 1×10-6 mbar | 10 mbar | – | 1 bar |
measurements | 5×10-9 mbar | – | – | – | – |
Simplified pumping and pressure measuring scheme (click for full version):

- Prec
- Succ >>
Ultima modifica il Giovedì, 19 Ottobre 2023 10:08