Detectors
Beamline Description | TARDI spectrometer | Focusing KB | Electromagnet endstation | RIXS endstation | Detectors | SLUANDOR IKON UHV camera (DO934P-BEN-9LH)
Andor open-front end iKon M SO CCD with DN63CF Camera for direct detection in vacuum applications
Sensor type: BEN, Back-illuminated CCD 1024x1024, enhanced
Pixels size: 13 μm
Image area: 13.3 x 13.3 mm with 100% fill factor
Active area pixel well depth (typical): 150.000 e-
Multiple Readout Rate: 5, 3, 1, 0.05 MHz
Read noise: 2.9 e- @ 0.05 MHz
Dark Current: 0.0001 e-/pixel/sec @ -100°C
Digitization: 16-bit @ all readout speed
Minimum temperature air cooled: -80ºC
Maximum cooling: -100°C
USB2 Interface
Integrated filter holder (filter not included)
Mounting: DN63CF flange (drawing attached to this offer)
Vacuum compatibility: > 10-8 mbar
![](/images/Documents/MAGNEDYN/Andor_Ikon.jpg)
Hamamatsu position sensitive MCP
Chevron assembly with phosphor screen output (F2225-21)
The phosphor screen coated on the glass plate converts the output charge distribution from the MCP into visibe image with a resolution of 80 um.
![](/images/Documents/MAGNEDYN/Hamamatsu.jpg)