SPELEEM HV conditioning
The following conditioning procedure must be followed only after venting and bake-out. Total time about 6 hours.
part 1: HV ramp up to 18kV:
NOTE: In the case of arcs (HV breakdowns), wait before incresing further the HV. Watch carefully the MCH and COL pressures and increase HV only when the gauge indicates that pressure is back to the normal value (chamber base pressure). |
part 2: energy analyzer rampingYou can now run the analyzer's electrodes simultanously:
|