Specifications
Beamline: insertion device
source: | Linear Planar Undulator (LPU) from Kyma s.r.l. divided in 2 sections |
period: | 46 mm |
number of periods: | 98 |
polarisation: | linear horizontal |
minimum gap: | 13.5 mm |
energy range: | 90 - 1200 eV |
peak brilliance: | ~ 4 x 1018 ph/s/100 mA/0.1% BW/mm2/mrad2 1) |
1) Evaluated for 400 eV photon energy with a storage ring energy of 2.0 GeV.
Harmonics at 2.0 GeV

SPEM microscope
|
toroidal prefocusing mirror |
Monocromator
Spherical Grating Monochromator (plane mirror and spherical grating) | |
fixed entrance slit (Hor) | |
fixed exit slit (Hor+Ver) | |
range: 400-1200 eV | |
resolving power: see Resolving Power of Monochromator |
Energy Resolution
Lateral Resolution
Sample
Samples must be conductive. Insulators are hardly imaged due to charging
Sample maximum dimension is 10x10x10 mm to fit sample Holder dimension
![]() |
Sample Holder scheme (pdf) |
Samples can be heated up during measurements to 1100 K by e-beam bombardment, filaments, Al2O3, BN heaters | |
Samples can be cooled down to 150 K using liquid nitrogen | |
Voltage and Current can be applied to the sample through electrical contacts (3 electrical contacts + 2 Termocouple contacts + GRD) |
|
The maximum pressure allowed during microscopy experiments is 1·10-6 mbar. |