Specifications
	
	Beamline: insertion device
| source: | Linear Planar Undulator (LPU) from Kyma s.r.l. divided in 2 sections | 
| period: | 46 mm | 
| number of periods: | 98 | 
| polarisation: | linear horizontal | 
| minimum gap: | 13.5 mm | 
| energy range: | 90 - 1200 eV | 
| peak brilliance: | ~ 4 x 1018 ph/s/100 mA/0.1% BW/mm2/mrad2 1) | 
1) Evaluated for 400 eV photon energy with a storage ring energy of 2.0 GeV.
Harmonics at 2.0 GeV

SPEM microscope
 
	
 | 
| toroidal prefocusing mirror | 
	
	
	
	Monocromator
| Spherical Grating Monochromator (plane mirror and spherical grating) | |
| fixed entrance slit (Hor) | |
| fixed exit slit (Hor+Ver) | |
| range: 400-1200 eV | |
| resolving power: see Resolving Power of Monochromator | 
Energy Resolution
	
Lateral Resolution
	
Sample
Samples must be conductive. Insulators are hardly imaged due to charging
Sample maximum dimension is 10x10x10 mm to fit sample Holder dimension
| 
				 | 
			Sample Holder scheme (pdf) | 
| Samples can be heated up during measurements to 1100 K by e-beam bombardment, filaments, Al2O3, BN heaters | |
| Samples can be cooled down to 150 K using liquid nitrogen | |
| 
				Voltage and Current can be applied to the sample through electrical contacts (3 electrical contacts + 2 Termocouple contacts + GRD)  | 
		|
| The maximum pressure allowed during microscopy experiments is 1·10-6 mbar. | 