Specifications
Pagina 4 di 5
Angular and Energy resolution
Energy resolution (currently the analyzer entrance slit size is 500 µm along energy dispersion and 12 mm along angle dispersion)
27 eV photons (5 o 10 meV line width)
| Pass Energy (V) | Resolution (meV) |
| 5 | 24 |
| 10 | 47 |
| 20 | 93 |
74 eV photons
| Pass Energy (V) |
Resolution (meV) monochromator slit 50 µm |
Resolution (meV) monochromator slit 100 µm |
| 5 | 50 | 90 |
| 10 | 65 | 100 |
| 20 | 102 | 130 |
Angular resolution
| Analyzer mode | Resolution along angular dispersion | Resolution along energy dispersion | Angle accepted along angular dispersion |
| WA | ±0.27o | ±0.67o | ±8o |
| AP0.75 | ±0.13o | ±0.33o | ±8o |
| AP1 | ±0.1o | ±0.25o | ±6o |
| AP1.5 | ±0.07o | ±0.17o | ±4o |
| AP2 | ±0.05o | ±0.13o | ±3o |
Ultima modifica il Giovedì, 08 Marzo 2012 14:49