Specifications
Pagina 2 di 5
Spatial resolution
The microprobe is formed using Schwartzschild objectives with multilayers. Two can be installed in UHV at the same time. We have available 27 eV, 74 eV and 95 eV objectives.
Flux and line width on the sample (monochromator slit 100 µm)
27 eV | 2x1010 photons sec -1 | 5 meV |
74 eV | 1.5x1011 photons sec -1 | 70 meV |
95 eV | 2x1011 photons sec -1 | 100 meV |
Long range background
![]() |
We use mostly 27 eV with less than 1 % Background and 74 eV with less than 25 % background. |
Microprobe size
![]() |
Focal depth
![]() |
Because of large (0.2) numeric apertire the focal depth is small. We need to focus with precision of approximately 2 µm. |
Ultima modifica il Giovedì, 08 Marzo 2012 14:49