The Endstations
Pagina 3 di 11
A: The Main Chamber
INSTRUMENT | STATUS/NOTES |
Electron analyzer: SCIENTA R3000. |
OK The Electron Analyser is at 60° from the incident photon beam, in the horizontal plane |
ComIXS x-ray fluorescence spectrometer; |
OK The Fluorescence Spectrometer is at -60° from the incident photon beam, in the horizontal plane |
Femptoamp meter for total yield x-ray absorption measurements (Keithley); Two (Keithley Electrometer 6430). |
OK Also two 6514 Keithley Electrometers and two Keithley Amplifier 428 are available. These instruments can be connected via PCI and/or GPIB interface with the PC that controls the entire beamline and they can collect signals coming from the mirror surfaces or a freshly evaporated gold mesh, the intensity read on some photodiodes, the drain current from the sample and the electron analyzer counts. |
Channeltron | OK |
Si photodiode | For Total Fluorescence Yield measurements only for very insulating samples |
VG Rear View with 8011 Electronics |
OK Works down to low energies (~15-25 eV). Retraction not working (LEED has been twisted) |
Retractable LEED (OCI) | Offline. It is currently in for repair. |
He lamp: SPECS UV-Source UVS 300 | Offline |
Manipulator (x,y,z and theta) (CREATEC) | OK |
LHe cryostat: from 30 K to about 1500 K
|
OK The minimum temperature on the sampe holder is 50 K using LHe, 130 K using LN2; |
Pressure gauge and Residual Gas Analyser | OK (p<3x10-10 mbar) |
Ar sputtering gun (VG) | OK. Only filament 2 is working. |
gas line for surface preparations in UHV; 2 leak valves |
O2, H2, CO are availabe. For other gases, please contact us |
single e-beam evaporator (Omicron) | Fe |
triple e-beam evaporator (Omicron) |
channel1: Si rod (OK? installed sep11) channel2:Mn basket (OK) channel3: Ti rod (filament maybe not working) |
sample scraping device | OK |
Mg/Al Kalpha X-Ray source |
Just installed (Sept 2011). To be tested working disatnce: 1-2 cm |
Fast entry Load Lock | OK |