Beamline
This software is used to control the beamline valves, read/set the energy and read the mesh current on the MSB-STATION computer.
Item | Description | Options |
Beam | beam status |
OK (beam available for users) not available |
Water |
cooling water status (prefocus, entrance slit, monochromator) |
OK (normal flux) warning (not enough flux) poor (no flux) |
Pressures | pressure status along the beamline |
OK (all safe) bad (not safe, e.g. during Ar sputtering) |
Valves |
status of 10 electropneumatic valves along the beamline |
all closed (safe) partly open (some open, some closed) in standby (valves 1-9 open, valve 10 closed; beamline warming up) all open (beam in the analysis chamber, ready for data acquisition) operating... (opening or closing in progress) |
Close | close all open valves | |
Standby |
set valves 1-9 open, valve 10 closed to warm up the beamline or keep warm after acquisition |
error is displayed if the opening is unsuccessful (e.g. interlocked by the beam operators during injection) |
Acquire |
open valve 10 to allow the beam in the analysis chamber for data acquisition |
|
Scheduling | schedule opening or closing the beamline |
OFF (no schedule) Close at [selected date/time], e.g. after the planned measurements Standby at [selected time], e.g. open valves 1-9 next morning to preheat the beamline before you arrive to work Standby when beam available preheat the beamline when the beam becomes available after a beam dump |
Energy | energy set by the monochromator | |
Set | tune the energy to the selected value | |
Mesh current | photocurrent from the gold mesh reflecting the photon flux | |
View last | time range of the horizontal axis of the mesh current graph | |
Cursor | position of the mouse pointer in the mesh current graph | |
EXIT | closes this software | |
Average mesh currents |
mesh currents at each energy averaged over the acquisition period; new averaging is started at each valve operation and energy tuning |
black values are valid red values are not valid (valves closed, energy tuning...) |