Twin anode X-ray source
Page 1 of 6
Sample Transfer and Measurement Positions
- Ensure the X-ray source is off and pulled out to the safe position marked by the top red line on the linear transfer.
|
X-ray source position for sample transfer |
![]() |
- Transfer the sample.
- Move the manipulator to the default position for XPS:
| Theta angle | 240° |
| x | 21.5 |
| y | 26.0 |
| z | 10.0 |
- Verify sample position in camera view. Orange circle indicates analyzed area (~7 mm).
| Analyzed area |
![]() |
- Adjust Z-axis (vertical) position.
-
Adjust horizontal position.
- X and Y must be moved almost simultaneosly. Change each single axis by no more than 1 mm.
- Per 1 mm of horizontal asjustment, move Y by 0.87 mm (sin(60°)) and X by -0.5 mm (-cos(60°)).
| Δd (mm) | X (mm) | Y (mm) | Δd (mm) | X | Y | |
| 0 | 21.5 | 26 | Default position | 0 | 21.5 | 26 |
| -1 | 21.0 | 26.9 | 1 | 22 | 25.1 | |
| -2 | 20.5 | 27.7 | 2 | 22.5 | 24.3 | |
| -3 | 20 | 28.6 | 3 | 23 | 23.4 | |
| -4 | 19.5 | 29.5 | 4 | 23.5 | 22.5 | |
| -5 | 19 | 30.3 | 5 | 24 | 21.7 |
- Lower X-ray gun to the red line on the linear transfer, approximately 12.5 mm from the bottom limit. (the width of wrench 13, which is often somewhere around.)
| X-ray gun operation position |
![]() |
- Cover windows 14, 15, 16, and 22 with lead shields.
| Lead shields |
![]() |
- Prev
- Next >>
Last Updated on Sunday, 06 July 2025 16:42