Safety
Safety protocols- ONLY FOR BACH staff
LHe DEWARS and REFILL
The dewar has a total of five ports:
(information taken from: http://www-ssrl.slac.stanford.edu/smbxas/instrumentation/cryoman.html#frost). |
Using Compressed GasUse only wrenches or tools provided by the cylinder supplier to remove a cylinder cap or to open a valve. Never use a screwdriver or pliers. Keep the cylinder valve closed except when in use. Attach the closed regulator to the cylinder. Never open the cylinder valve unless the regulator is completely closed. Regulators are specific to the gas involved. A regulator should be attached to a cylinder without forcing the threads. Ensure the threads of both the regulator and main valve are clean. If the inlet of a regulator does not fit the cylinder outlet, no effort should be made to try to force the fitting. A poor fit may indicate that the regulator is not intended for use on the gas chosen. Turn the delivery pressure adjusting screw counter-clockwise until it turns freely. This prevents unintended gas flow into the regulator. Open the cylinder slowly until the inlet gauge on the regulator registers the cylinder pressure. If the cylinder pressure reading is lower than expected, the cylinder valve may be leaking. With the flow control valve at the regulator outlet closed, turn the delivery pressure adjusting screw clockwise until the required delivery pressure is reached. Check for leaks using Snoop or soap solution. At or below freezing temperatures, use a glycerin and water solution, such as Snoop, rather than soap. Never use an open flame to detect leaks. When finished with the gas, close the cylinder valve, release the regulator pressure and replace the gas cap if it will not be used in the near future.
|
Procedure to access the FE hutch and radiation emergency |
Magnetic fields and cryomagnet endstation Hazard and SafetyFirst we remind that it is forbidden for the users to access the high platform around the cryomagnet and that it is forbidden to leave the beamline unattended when the magnetic field is on.
Special care has to be observed when operating with the superconducting coil since high magnetic field is generated in the area around the cryomagnet. Exposure should be always avoided for pregnant women, people with cardiac pacemaker, neurostimulators, cochlear implant, derivation valves, lenses, intra-corporeal metallic foreign objects. |
In this table we report the magnetic field measured along the field at 314 mm from the center of the coil (maximum field in the restricted area external to the cryomagnet):
|
▪ Projectile effect (attraction by a static magnetic field and acceleration, with speeds of up to several meters per second): ferromagnetic material (if in doubt about the ferromagnetic nature of a metal object, a test can be carried out using a small magnet). REMOVE all metallic objects which could be attracted by the magnetic field (screwdrivers, screws, bolts, nuts, ....) from the magnetic field area!! Always be aware that even a small object can become a projectile and cause fatal injuries.Due to the presence of a strong magnetic field, certain materials may present a functional or even a vital risk: ▪ Displacement of intra-corporeal metallic foreign objects: Intraocular metallic foreign body (metal worker, history of ballistic orbit trauma, old intra-cranial aneurysm clips) ▪ Perturbed functioning of certain devices: cardiac pacemaker, neurostimulators, cochlear implant, derivation valves, lenses ▪ Credit cards, mobile self phones, watches, tape recorders and cameras, magnetic cards can be permanently damaged and information contained could be corrupted and should stay outside the magnetic field area)
• Electrical transformers may become magnetically saturated in fields above 50 gauss. The safety characteristics of equipment may also be affected. □ I have read and accept
|