Scienta R3000: non-linear intensity scale
Page 1 of 2
R3000 XPS Intensity scale calibration
Note: All calibration curves were done with detector area set to 640x384 px2, where 640px corresponds to full x (energy) scale, 384px to Y-(angular) scale between 53 and 436px.
Fixed Mode (Calibration → Voltages...→ Area)


Swept Mode


Graphs in higher resolution.
Approximate relation between intensity in swept and fixed mode (full detector range):
SES, Igor Pro:
Cps = C / n/ StepTime
where,
Cps = couts per second in fixed mode ( approximately what we see in Calibration → Voltages →Area)
C= counts acquired in swept angular mode
n = number of sweeps