Manual_Annealing
Page 2 of 2
PBN sample carriers
IMPORTANT
When the annealing on PBN sample holders is done in the preparation chamber or load-lock, the true temperatures are approximately 150°C higher than those in the following tables for the Main chamber where the sample carrieres are in good thermal contact with thy cryostat.
"Old" PBN sample holder
- Manipulator cooled by air flow
Curret (A) | Voltage (V) | Temperature (°C) |
0.4 | 6 | 100°C |
0.6 | 8 | 150°C |
0.7-0.8 | 10 | 200°C |
1.1 | 13.6 | 300°C |
1.2 | 15 | 327°C |
1.6 | 18.1 | 400°C |
2.1 | 22.4 | 500°C |
"New" PBN sample holder
- Manipulator cooled by air flow
Curret (A) | Voltage (V) | Temperature (°C) |
0.5 | 7.5 | 100°C |
0.8 | 11.3 | 200°C |
1.3 | 15.3 | 300°C |
1.7 | 18.7 | 400°C |
2.1 | 22.2 | 500°C |
"All metal" PBN sample holder
- Manipulator cooled by air flowCurret (A) | Voltage (V) | Temperature (°C) |
1 | 12.2 | 270°C |
1.2 | 14.5 | 327°C |
1.8 | 19.4 | 470°C |
2.0 | 21.4 | 504°C |
"Elena" PBN sample holder
- Manipulator cooled by air flowCurret (A) | Voltage (V) | Temperature (°C) |
1.0-1.1 | 9.2 | 256°C |
1.8-1.9 | 14.4 | 420°C |
Large E-beam sample carrier
Filament radiation annealing, manipulator cooled by air flow
Technical Note: Thoriated W filament, 0.2mm, 13-14 threadsCurrent (A) | Voltage (V) | Power (W) | Temperature(°C) |
1.4 | 1.8 | 2.5 | 50°C |
1.8 | 2.5 | 4.5 | 100°C |
2.5 | 4.4 | 11 | 210°C |
2.9 | 5.4 | 15.7 | 265°C |
3.0 | 5.7 | 17.1 | 283°C |
3.2 | 6.3 | 20.2 | 300°C |
3.5 | 7.4 | 25.9 | 350°C |
3.6 | 7.5 | 27 | 361°C |
3.8 | 8.3 | 31.5 | 388°C |
3.9 | 8.6 | 33.5 | 402°C |
Estimated sample temperature vs heater current
HV E-beam annealing, manipulator cooled by air flow
Warning 1: HV on thermocouple pins!Warning 2: Operation allowed ontly to the staff.
Warning 3: Filament must be grounded.
Warning 4: Never use PID box to monitor the temperature.
Note: Following data (collected in November 2018) are only approximate, they depend on the filament shape and distance from the Mo plate; the pressure and nature of the used gas.
Filament (A) | Emission (mA) |
Pyrometer (°C) Emis.=16 |
Temperature (°C) +- 30°C |
Comment |
2.3 | 30 | 420 | 506 | |
2.4 | 35 | 550 | 623 | |
3.8 | 40-42.3 | 570-600 | 638-665 | pO2=5x10-7 mbar |
3.8 | 51.2 | 660 | 736 | |
3.8 | 57.4 | 690 | 760 | pN3B3H6=5x10-7mbar |
Small E-beam sample carrier
Filament radiation annealing, manipulator cooled by air flow
Technical Note: Thoriated W filament, 0.2mm, 13-14 threadsCurrent (A) | Voltage (V) | Power (W) | Temperature(°C) | Comment |
1.4 | 1.8 | 2.5 | 102 | After >10 minutes |
3.2 | 6.2 | 20 | 430 +- 30 °C |
HV E-beam annealing, manipulator cooled by air flow
Warning 1: HV on thermocouple pins!Warning 2: Operation allowed ontly to the staff.
Warning 3: Filament must be grounded.
Warning 4: Never use PID box to monitor the temperature.
Note: December 2018, HV switched on when Filament current is below 1.5A. Pyrometer aimed on the Mo plate close to the crystal sample.
Filament (A) | Emission (mA) |
Pyrometer (°C) Emis.=16 |
Temperature (°C) +- 30°C |
Comment |
3.7 | 6 | 440°C | 500°C | pO2=2x10-6 mbar |
3.8 | 13.2 | 548°C | 600°C | pO2=2x10-6 mbar |